EQUIPMENT & TECHNOLOGY

CALIBRATION LABORATORY

Our calibration laboratory is focused on calibration of hygrometers for both trace moisture (dew point) measurement and measurement of relative humidity. We also able to calibrate thermometers,  usually as a part of instruments for humidity measurement.

CALIBRATION RANGES

  • DEW POINT: -80 to + 20 °C DP or -100 to + 25 °C DP
  • RELATIVE HUMIDITY: 0 to 100 % RH
  • TEMPERATURE: -40 to +100 °C 

INSTRUMENTS USED FOR CALIBRATION IN THE TRACE MOISTURE RANGE

  • DG-2S DEWPOINT GENERATOR (Michell Instruments Ltd)
  • DEWPOINT HYGROMETER Series 4000 (Michell Instruments Ltd)
  • DEWPOINT HYGROMETER Series 3000 (Michell Instruments Ltd)
  • Moisture analyzer SYSTEM 1112-0222001 (SENSORIKA)

INSTRUMENTS USED FOR CALIBRATION IN THE RELATIVE HUMITIDY RANGE

  • C-1 Relative Humidity Generator (GENERAL EASTERN)
  • HYGROMETER HG 120 with a probe HTP-2714220 (SENSORIKA)

 INSTRUMENTS USED FOR CALIBRATION OF THERMOMETERS

  • Laboratory resistance thermometer Pt100 type LW 3  (Heraeus)
  • Laboratory thermometer EXT-809-97 (EXATHERM)

Our calibration laboratory specializes on calibration of SENSORIKA HUMISTARTM measuring instruments. Instruments from other manufacturers can be calibrated on demand.

Our calibration equipment is traceable to British (NPL) and American (NIST) national humidity standards. Traceability of the equipment is declared in the document below.